Thursday, October 3, 2013

EEG and Other Technologies

MEG VS EEG

Today’s MEG commercial systems are organized in whole-head sensor arrays arranged in a rigid helmet covering most of the head surface but the face area. MEG signals are recorded from about 300 channels, which sometimes consist of pairs of magnetometers to form physical gradiometers (Hämäläinen, Hari, Ilmoniemi, Knuutila, & Lounasmaa, 1993). These latter are less sensitive to far-field sources, which are supposed to originate from distant generators (e.g., road traffic, elevators, heartbeats). An important benefit of MEG systems is the possibility to record EEG from dense arrays of electrodes (>60) simultaneously, thereby completing the electromagnetic signature of neural currents.
Additional analog channels are usually available for miscellaneous recordings (heart monitoring (ECG), muscle activity (EMG), eye movements (EOG), respiration, skin conductance, subject’s responses, etc.). Sampling rate can reach up to 5KHz on all channels with a typical instrumental noise level limited to a few fT/sqrt(Hz). One femto-Tesla (1fT) is 10-15T. Ongoing brain signals measured with MEG are on the range of about [10,50] fT/sqrt(Hz), with a relatively rapid decay in amplitude as frequency increases.


MEG has substantial benefits with respect to EEG:

  1. while EEG is strongly degraded by the heterogeneity in conductivity within head tissues (e.g., insulating skull vs. conducting scalp), this effect is extremely limited in MEG, resulting in greater spatial discrimination of neural contributions. This has important implications for source modeling as we shall see below;
  2. Subject preparation time is reduced considerably;
  3. Measures are absolute, i.e. they are not dependent on the choice of a reference;
  4. Subject’s comfort is improved as there is no direct contact of the sensors on the skin.
MEG/EEG experiments can be run with the subjects in supine or seated positions. A caveat however concerns EEG recording in supine position, which may rapidly lead to subject discomfort because occipital electrodes become painful pressure points. The quiet, room-size and fairly open environment of the MSR and Faraday cages (relatively to MRI bores), make it more friendly to most subjects. Care givers may accompany subjects during the experiment.
Installation of new MEG systems is presently steadily growing within research and clinical centers (about 200 worldwide).



On the benefits of a larger number of sensors: (a) 3D rendering of a subject’s scalp surface with crosshair markers representing the locations of 151 axial gradiometers as MEG sensors (coil locations are from the VSM MedTech 151 Omega System). (b) Interpolated field topography onto the scalp surface 50 ms following the electric stimulation of the right index finger. The fields reveal a strong and focal dipolar structure above the contralateral central cortex. (c) The number of channels has been evenly reduced to 27. Though the dipolar pattern is still detected, its spatial extension is more smeared – hence the intrinsic spatial resolution of the measurements has been degraded – due to the effect of interpolation between sensors, which are now quite distant from the maxima of the evoked magnetic field pattern.

Introduction to EEG and MEG


Basics of EEG and MEG: Physiology and data analysis

1. Introduction


The electrical activity of active nerve cells in the brain produces currents spreading through the head. These currents also reach the scalp surface, and resulting voltage differences on the scalp can be recorded as the electroencephalogram (EEG). The currents inside the head produce magnetic fields which can be measured above the scalp surface as the magnetoencephalogram (MEG). EEG and MEG reflect brain electrical activity with millisecond temporal resolution, and are the most direct correlate of on-line brain processing obtainable non-invasively. Unfortunately, the spatial resolution of these methods is limited for principal physical reasons. Even with an infinite amount of EEG and MEG recordings around the head, a non-ambiguous localisation of the activity inside the brain would not be possible. This "inverse problem" is comparable to reconstructing an object from its shadow: only some features (the shape) are uniquely determined, others have to be deduced on the ground of additional information. However, by imposing reasonable modelling constraints or by focussing on rough features of the activity distribution, useful inferences about the activity of interest can be made.
The continuous or spontaneous EEG (i.e. the signal which can be viewed directly during the recording) can be helpful in clinical environments, e.g. for diagnosing epilepsy or tumours, predicting epileptic seizures, detecting abnormal brain states or classifying sleep stages. The investigation of more specific perceptual or cognitive processes requires more sophisticated data processing, like averaging the signal over many trials. The resulting "event-related potentials" or "event-related magnetic fields" are characterised by a series of deflections or "components" in their time course, which are differentially pronounced at different recording sites on the head. They are classically labelled according to their polarity (positive/negative) at specific recording sites and the typical latency of their occurrence (e.g. N100 refers to a negative potential around 100ms, similarly for P300, N400 etc.). The amplitude and pattern of these components can be interpreted as dependent variables, which are distinguished by their dependency on the task, stimulus parameters, degree of attention etc.. Components related to basic perceptual processing (somato-sensory, visual, auditory, which are also termed "evoked potentials" or "evoked magnetic fields") can be used by clinicians to check the connection of the periphery to the cortex, or can assist in the diagnosis of brain death. Event-related potentials and magnetic fields are widely used to characterise the spatio-temporal pattern of brain activity in basically all areas of neuroscience. Latest developments include the combination of EEG/MEG with functional magnetic resonance imaging, for combining the superior spatial resolution of the latter with the better temporal resolution of the former.

2. Generation of the signals


If an electrical signal is transmitted along an axon or dendrite, electrical charges are separated along very short distances on the corresponding cell membranes. Separated charges act as a small temporary "battery", with positive polarity at the side where the current is leaving and with negative polarity where the current is returning. These "batteries" are called "primary currents" since they are the sources of interest that reflect directly the activity of the corresponding neurons. The primary currents are embedded in a conductive medium, i.e. the brain tissue and brain liquor. Therefore a current is induced which flows through these media, as well as the skull and scalp, called "secondary" or "volume" currents. EEG would not be possible without volume currents, which also reach the scalp surface and cause voltage differences at the scalp that can be picked up by EEG electrodes. Both primary and volume currents produce magnetic fields, which sum up and can be measured by pick-up coils above the head using MEG.
Possible contributors to the measurable signals are: 1) Action potentials along the axons connecting neurons, 2) currents through the synaptic clefts connecting axons with neurons/dendrites, and 3) currents along dendrites from synapses to the soma of neurons. Action potentials are "quadrupolar", i.e. two opposing currents occur in vicinity of each other, the effects of which cancel each other out. In addition, action potentials last only a few milliseconds, making it more difficult to achieve a larger number of them being active simultaneously and to sum up to a measurable signal at larger distance. Synapses are very small and the current flow relatively slow. Furthermore, they can be located nearly randomly around a neuron and its dendrites, so that the contributions of different synapses are likely to cancel each other out. The current flow along post-synaptical dendrites is dipolar, and at least the apical dendrites in cortical pyramidal cells are generally parallely organised. They are typically active for about 10ms after synaptic input. For these reasons, the apical dendrites of the cortex are assumed to contribute strongest to the measurable EEG and MEG signals. Because this activity directly reflects processing of specific neurons, EEG and MEG are the most direct correlate of on-line brain processing obtainable non-invasively. However, one dendrite would be far too weak to produce a measurable signal, instead tens of thousands are required to be active synchronously. Consequently, EEG/MEG are only sensitive to coherent simultaneous activity of a large number of neurons. For modelling purposes it is important to note that many weak neighbouring dipolar sources can be summarised to one "dipole", which can be interpreted as a "piece of active cortex".
Dipolephysiology.gif

3. General data analysis


Sampling Recordings are taken virtually simultaneously at all recording sites at different time points. The time difference between sampling points is called the sampling interval. The sampling rate in Hertz (Hz) is given by the number of samples per second, i.e. (1000ms/{sampling interval in ms}). Typical sample intervals are of the order of several milliseconds, depending on the purpose of the study. If very short-lived brain processes are of interest, i.e. if the signal is to contain high frequency components, a higher sampling rate is required (for early somato-sensory or brain stem potentials, for example, a sampling rate of about 1000Hz or more is required). This, however, also means more data, i.e. more storage space is required and further processing will be slowed down. If slower brain processes are studied and only low frequency components are of interest, the sampling rate can be reduced (e.g. about 40Hz for slow waves like the readiness potential). In typical event-related potential or field studies, where the interesting frequencies usually lie between 0-100 Hz, a sampling rate of 200-500 Hz is common. With modern analysis equipment, the computing power and storage space is usually not a crucial issue any more.
Filtering If the signal of interest (e.g. a specific ERP component) is already roughly known, filters can be applied that suppress noise in frequency ranges where the signal has low amplitude. High-pass filters attenuate frequencies below a certain cut-off frequency (i.e. they "let pass" frequencies above it), while low-pass filters behave the other way round. A combination of a low-pass and a high-pass filter is called a band-pass filter. A typical band-pass filter for ERP studies would be 0.1-30Hz, but this highly depends on the signal of interest and the purpose of the study.
Averaging The brain response of a single event (i.e. the signal after the presentation of a single stimulus) is usually too weak to be detectable. The technique usually employed to cope with this problem is to average over many similar trials. The brain response following a certain stimulus in a certain task is assumed to be the same or at least very similar from trial to trial. That means the assumption is made that the brain response does not considerably change its timing or spatial distribution during the experiment. The data are therefore divided in time segments (or "bins") of a fixed length (e.g. one second) where the time point zero is defined as the onset of the stimulus, for example. These time segments can then be averaged together, either across all stimuli present in the study, or for sub-groups of stimuli that shall be compared to each other. Doing so, any random fluctuations will cancel each other out, since they might be positive in one segment, but negative in another. In contrast, any brain response time-locked to the presentation of the stimulus will add up constructively, and finally be visible in the average. The noise level is typically about 20uV, and the signal of interest about 5uV. If the noise was completely random, to reduce the amplitude of the noise by a factor of n one would have to average across n*n samples. Averaging over 100 segments would therefore reduce the noise level to about 2uV, at least less than half the amplitude of the signal of interest. A further 100 segments (i.e. 200 segments altogether) would reduce the noise level to about 1.4uV. To reduce the noise level achieved by 100 segments by a factor 2, one would need 400 segments (20uV/20=1uV), i.e. 4 times as many. The fact that the noise level in the average data decreases only with the square root of the number of trials poses a considerable limitation on the duration of an experiment, and should be carefully considered prior to setting up a study.
Baseline correction During the recording, EEG and MEG signals are undergoing slow shifts over time, such that the zero level might differ considerably across channels, though the smaller and faster signal fluctuation are similar. These signal shifts can be due to brain activity, but can also be caused by sweating (in the case of EEG), muscle tension, or other noise sources. It would therefore be desirable to have a time range where one can reasonably assume that the brain is not producing any stimulus related activity, and that any shift from the zero line is likely due to noise. In most ERP and ERF studies, this "baseline interval" is defined as several tens or hundreds of milliseconds preceding the stimulus. For each recording channel, the mean signal over this interval is computed, and subtracted from the signal at all time points. This procedure is usually referred to as "baseline correction". It is crucial in ERP and ERP experiments to ensure that an observed effect (e.g. an amplitude difference between the components evoked by two sorts of stimuli) is not already present in the signal before the stimuli were actually present. If this was the case, it would indicate that the signal is contaminated by a confound that is not stimulus-related.
Display In usual displays for time courses of EEG and MEG data, time is chosen as the x-axis and the amplitude as the y-axis. Traditionally, negative amplitudes are often (but not consistently) plotted upwards for EEG data. If multi-channel ERP and ERF recordings are presented, it is usual to project the recordings sites on a two-dimensional surface, and to plot the curves for these recording sites at the corresponding location. This gives a rough impression of the spatial distribution of the signal over time. However, with increasing number of recording channels, these images become very difficult to interpret. If the spatial distribution of a specific time point or time range (e.g. a peak) in the signal shall be analysed in more detail, the signal can be interpolated between recording sites and plotted as a colour-coded map. Often positive amplitudes are displayed in red colour, and negative amplitudes by blue colour.
EEGpreprocessing.gif

4. EEG recordings


Referencing Electric potentials are only defined with respect to a reference, i.e. an arbitrarily chosen "zero level". The choice of the reference may differ depending on the purpose of the recording. This is similar to measures of height, where the zero level can be at sea level for the height of mountains, or at ground level for the height of a building, for example. For each EEG recording, a "reference electrode" has to be selected in advance. Ideally, this electrode would be affected by global voltage changes in the same manner as all the other electrodes, such that brain unspecific activity is subtracted out by the referencing (e.g. slow voltage shifts due to sweating). Also, the reference should not pick up signals which are not intended to be recorded, like heart activity, which would be "subtracted in" by the referencing. In most studies, a reference on the head but at some distance from the other recording electrodes is chosen. Such a reference can be the ear-lobes, the nose, or the mastoids (i.e. the bone behind the ears). With multi-channel recordings (e.g. >32 channels), it is common to compute the "average reference", i.e. to subtract the average over all electrodes from each electrodes for each time point. This distributes the "responsibility" over all electrodes, rather than assigning it to only one of them. If a single reference electrode was used during the recording, it is always possible to re-reference the data to any of the recording electrodes (or combinations of them, like their average) at a later stage of processing. In some cases "bipolar" recordings are carried out, where electrode pairs are applied and referenced against each other for each pair (e.g. left-right symmetrical electrodes).
Electrode locations To compare results over different studies, electrode locations must be standardised. These locations should be easily determinable for individual subjects. Common is the extended 10/20 system, where electrode locations are defined with respect to fractions of the distance between nasion-inion (front-back) and the pre-auricular points (left-right). Obviously, the relative locations of the electrodes with respect to specific brain structures can only be estimated very roughly. If this is a crucial point, electrode locations can be digitised together with some anatomical landmarks (like the pre-auricular points, the ear-holes, nasion or inion). These landmarks can also be detected in individual MRI or CT scans if available. The landmarks of the different co-ordinate systems (EEG digitisation, MRI/CT scan) can then be matched, such that both data sets are in the same co-ordinate system. Another possibility is to digitise a large number of points (at least several hundreds) evenly distributed over the scalp surface, and match those to the scalp surface segmented from the individual MRI/CT image. This makes possible more realistic modelling of the head geometry for sophisticated source estimation, and furthermore the visualisation of the results with respect to the individual brain geometry.
Fixation of electrodes Clinical recordings usually employ only few (1-9) electrodes at positions standardised for given purposes. Modern ERP studies that intend topographic analysis often employ 32 to 256 electrodes. If only a few electrodes are used, these can be glued to the skin by hand, using electrically conducting glue. With larger numbers of electrodes, electrodes are commonly mounted on a cap. Contact with the skin has then to be made by introducing a electrically conducting substance between skin and electrode. Depending on the device, this can be conducting gel, glue, or sponges soaked with a saline solution. For special applications, needle electrodes (subduran) are used. Electrodes can be of different materials: Tin electrodes are cheap but not recommended at low frequencies. Ag/AgCl (silver-silver-chloride) electrodes are suitable for a larger frequency range and most commonly used in research. Gold and platin electrodes are used for special requirements. The amplitude of the raw signal that is recorded is of the order of 50 uV, i.e. about a million times weaker than those by household batteries. It therefore required special amplifiers that amplify the signal and convert it to a digital signal that can be stored and processed on a computer. This has to be accomplished at high sampling rates (several hundred or thousand recordings a second), and virtually simultaneously for each recording channel.

5. MEG recordings


SQUIDS The amplitude of the magnetic fields produced spontaneously by the human brain are of the order piko-Tesla when recorded, and that of the evoked responses is around 100fT (i.e. more than one million times smaller than the earth's magnetic field). These very low field strenghts can be recorded by so-called "Superconducting Interference Devices" (SQUIDS): A specially prepared pick-up coil is cooled down to superconductivity by liquid helium, and exploiting a specific quantum effect the magnetic flux through the coil is measured.
Influence of coil types The magnetic field, in contrast to the electric potential, has a direction, usually illustrated by magnetic field lines. A current flowing along a straight line produces circular magnetic field lines that are concentric with respect to the current line. If the thumb of your right hand points in the direction of the current flow (which is from negative to positive polarity, or "from minus to plus"), then the remaining fingers point in the direction of the magnetic field ('right-hand-rule'). The coils pick up only the strength of the magnetic field in the direction perpendicular to the coil area. Different coil configurations are in use: A single coil is called a 'magnetometer'. If the difference between two or more neighbouring coils is measured, it is called a (first order, second order etc.) 'gradiometer'. Gradiometers have the property to be less sensitive to sources at larger distances, because those have more similar field strengths at all coils, which in the difference yields lower values. This has the advantage that noise sources at larger distances are also suppressed (like the heart, or electric devices near the recording system). Because MEG sensors are so very sensitive, they have to be well-shielded against magnetic noise. The shielding of the laboratory and the SQUID technology make an MEG device 10 to 100 times more expensive than an EEG system, and the permanent helium-cooling imposes considerable maintenance costs.
Insensitivity to radial and deep sources A special and important feature of MEG is its insensitivity to current sources which are "radial", i.e. which are directed towards or away from the scalp (like the top of a cortical gyrus). It mainly "sees" tangential sources, which are parallel to the scalp. The full explanation for this characteristic would involve a fair amount of physics and mathematics, but an idea can given for the following simplified situation. Imagine the head was modelled as a sphere, and the recording coil would be tangentially placed above its surface. A dipole within the sphere pointing towards the coil would produce a magnetic field parallel to the coil, which would therefore not be measured. The volume currents, on the other side, are symmetric with respect to the axis along the dipole direction. This means that the magnetic field through the coil by the current on one side of this axis is cancelled by the current on the other side. Therefore the contribution of the primary current as well as that of the volume current vanish. More detailed analysis of this problem shows that the magnetic field of a radial dipole in a spherical volume conductor generally vanishes at all possible recording sites. A special case of a radial dipole is a dipole located at the centre of the sphere: It must necessarily point away from the centre, and is therefore radial and "invisible" for MEG. Roughly speaking, the nearer a source is to the centre, "the more radial" it is. This implies that the deeper a source, the less visible it is for MEG. This seems to be a disadvantage for the MEG, but turns into an advantage if superficial and tangential sources are targeted at, since in that case the corresponding MEG signal is less contaminated by other disturbing sources.

6. Basic differences between EEG and MEG


Both EEG and MEG are sensitive to the topography of the brain – the details of the structure of its surface. The brain surface is quite dramatically folded, and whether the signal arises in the convex tops of the folds (gyri) or the concave depths of the folds (sulci) affects sensitivity to neuronal current sources (dipoles). MEG is particularly insensitive to radial dipoles, those directed towards or away from the scalp (like at the top of a gyrus). It mainly 'sees' tangential dipoles, which are parallel to the scalp. The reason for this is that if the head were a perfect sphere (a reasonable first approximation for an average head shape), all magnetic fields resulting from a radial dipole (including the volume currents) cancel each other out at the surface (a more detailed explanation would involve a fair amounts of physics). EEG is sensitive to both radial and tangential dipoles.
A special case of a radial dipole is a dipole located at the centre of a sphere: it must necessarily point away from the centre, and is therefore radial and 'invisible' for MEG. Roughly speaking, the nearer a source is to the centre, 'the more radial' it is. This implies that the deeper a source is in the head, the less visible it is for MEG. This seems to be a disadvantage for MEG, but turns into an advantage if superficial and tangential sources (e.g. in sulci of the cortex) are targeted, since in that case the corresponding MEG signal is less contaminated than EEG by other possibly disturbing sources. Furthermore, it has been shown that superficial and tangential sources can be localized with more accuracy than with EEG, particular in somatosensory and some auditory parts of the brain. MEG has therefore been widely used to study early acoustic and phonetic processing, and the plasticity of the somatosensory and auditory system.
The basics of EEG were invented in the 1920s, mainly with medical applications in mind. EEG is widely used in clinical diagnostics and research. The first MEG systems (starting with 1-channel!) appeared in the late 1960s, and first commercial systems took off in the 1980. The development was mainly driven by engineers and physicists. This, and possibly the fact that MEG is much more expensive than EEG, often makes these two look like very different methods with very different applications. It is important to keep in mind, however, that both techniques measure the same type of brain activity from slightly different angles. As a general rule, EEG "sees more" of the brain, but is less able to localise this activity. MEG "sees less" of the brain, but what it sees it can localise better. As we will see below, a combination of EEG and MEG is superior for localisation than each method on its own. If localisation is not the main focus of a study, EEG may often be sufficient, for example in many clinical environments.

7. Evoked and Induced Brain Responses


The averaging procedure described above assumes that there is something like the average brain response, i.e. a signal with roughly the same time course in every epoch for a particular stimulus category. But what if the timing changes slightly from epoch to epoch? If this jitter across epochs is small compared to the time-scale of the signal, one does not have to worry. But what if the signal consists of a rapid sequence of positive and negative deflections (e.g. an "oscillation")? Then the jitter may sometimes result in a positive value in some epochs, and negative values in other epochs. Across epochs, this may average out to zero. This could be the case with co-called "gamma oscillations" – a signal in the frequency range above 30 Hz (definitions differ slightly), which has been associated with complex stimulus processing such as Gestalt perception. These gamma oscillations may occur time-locked to a stimulus (e.g. always between 200 and 300 ms after picture onset). However, the "phase information", i.e. the exact occurrence in time of maxima and minima of the oscillations, may vary across epochs. These signals are called "induced", or order to distinguish them from the above-mentioned "evoked" signals that are both time- and phase-locked to the stimulus. In order to extract induced signals from the data, the frequency power spectrum (i.e. the signal amplitude for different frequency bands) has to be computed for every epoch separately. The power spectrum only contains information about the amplitude of the signal for a certain frequency in a certain latency range, but no phase information. If there are consistent gamma oscillations between 200 and 300 ms after every picture, the spectral power in the gamma frequency range should be consistently above baseline level in the latency range across epochs. This method has been applied to studies on Gestalt perception, as illustrated in the figure below. For example, it has been found that gamma band activity is usually larger when an incomplete but meaningful pattern is perceived, compared to either non-meaningful or easily recognisable patterns.
TimeFrequency.jpg

8. Statistical analysis


The statistical analysis of EEG/MEG data is as yet not fully standardised, and the strategies employed can vary considerably depending on the hypothesis and purpose underlying the experiment. The most frequent approach is the "Analysis of Variance" or "ANOVA". The signal amplitudes for different electrodes, sensors or dipoles are considered as dependent variables, and extracted for each subject (e.g. like mean reaction times or error rates per subject in behavioural experiments). Imagine a so-called oddball experiment, in which subjects are presented with two different tones: One tone of 600Hz, which occurs 90% of the time, and another tone of 1000Hz, which occurs only 10% of the time. Subjects are required to press different buttons in response to the frequent and rare tones. Does the maximum amplitude of the ERP in a certain time range change as a function of the probability of the tone? In this case, one would first have to decide which time range to analyse. In this kind of oddball experiment, it is known that rare tones produce a so-called "P300", i.e. a positive deflection of the ERP around centro-parietal electrode sites, which usually reaches its maximum around 300ms after stimulus onset. It would therefore be reasonable to look for the largest peak around 300ms at the electrode Pz, and to extract the amplitudes at this electrode and this point in time for all subjects. These values would enter the ANOVA as a dependent variable. "Probability" would be a single, two-level factor allowing for a comparison between values obtained for the presentation of the rare tones with those obtained for the frequent tones. In general, the result is that the amplitude of the P300 is larger for rare tones. Does only the amplitude change with the toneâs probability of occurrence, or could there also be a change of "Laterality"? In other words, could the rare tones produce larger amplitudes in either the left or right hemisphere compared with the frequent ones? In this case, one would have to extract amplitude values for at least two electrodes, one in the left and one in the right hemisphere (e.g. P5 and P6, left and right of Pz). The ANOVA now has two factors: Probability (rare and frequent) and Laterality (left and right). If the laterality changes with tone probability, one should obtain an interaction between Probability and Laterality. In the case of simple tones, such an interaction is generally not found, although it might be expected for more complex (e.g. language-related) stimuli. If such a hypothesis-based approach is not suitable (e.g. because there is no specific hypothesis, or because the hypothesis was falsified but one still does not want to give up...), exploratory strategies have to be used. For example, separate statistical tests (like paired T-tests) can be computed electrode by electrode, or dipole by dipole. However, if we perform statistical comparisons at different recording sites or source locations, and each of those by itself has a chance of 5% to produce a false positive (i.e. tells us there is a significant difference when in fact there is not), then we have to expect that a fraction of 5% of our locations are incorrectly labeled as significant. This is the common problem of multiple comparisons. Methods to deal with this problem in the case of EEG and MEG data are currently under development.

9. Estimating the neuronal sources


Underdeterminacy of the inverse problem Even if the EEG and MEG were measured simultaneously at infinitely many points around the head, the information would still be insufficient to uniquely compute the distribution of currents within the brain that generated these signals. The so-called "inverse problem", i.e. the estimation of the sources given the measured signals, is "under-determined" or "ill-posed". But all is not lost: By imposing appropriate modelling constraints, one can often derive valuable information about spatial features of the source distribution from the data. The main questions in this context are: Do I have enough information from other sources (neuropsychology, other neuro-imaging results) to narrow down the possible generators? Or can I ask my questions such that it is not necessary to know the exact source distribution, but only some rough features that can be determined without further modelling constraints?
Approaches to deal with the inverse problem One can distinguish between two strategies to tackle the inverse problem: 1) Approaches that make specific modelling assumptions, like the number of focal sources needed to produce the recorded data and their approximate location, and 2) approaches that make as little modelling assumptions as possible, and focus on rough features of the current distribution that are determined by the data rather than specific modelling assumptions (like laterality). Under 1) fall the so-called 'dipole models': The number of active dipoles, i.e. small pieces of activated cortex, is assumed to be known (usually only a few). An initial guess about their location and orientation is made, and these parameters are then adjusted step by step until the predicted electric potential or magnetic field resembles the measured one within certain limits. These methods have been successfully applied in studies of perceptual processes and early evoked components. Point 2) comprises the 'distributed source models': The naturally continuous real source distribution is approximated by a large number (several hundreds or thousands) of sources equally distributed over the brain. The strengths of all these sources cannot be estimated independently of each other, but at least a "blurred" version of the real source distribution can often be obtained. Peaks in this estimated current distribution might correspond to centres of activity in the real current distribution. These methods are preferred if modelling assumptions for dipole models cannot be justified, like in cognitive tasks or in noisy data.
Sourceestimation.gif

10. EEG, MEG and fMRI: Localisation Issues


Although the signals of EEG and MEG are generated by the same sources (electrical currents in the brain), they are both sensitive to different aspects of these sources (as described above). This could be compared to viewing the shadows of the same object from two different angles. It should not be surprising then that the information of EEG and MEG are complementary, and combining the two in source estimation usually yields more accurate result than using either method alone. However, this also means more work, and not everybody has both EEG and MEG available. It is therefore important to decide at the beginning of a study 1) whether localisation is required at all; 2) whether absolute anatomical localisation is required (e.g. in combination with structural MRI images), or whether relative localisation (e.g. "left more than right") is sufficient; 3) whether it is already known that the sources are visible in EEG, MEG or both.
While for EEG and MEG, we can safely assume that we are measuring the same sources from different angles, the same cannot be said for fMRI compared to EEG/MEG. Standard fMRI measures a blood-oxygen-level dependent (BOLD) signal. This is a metabolic (rather than electric) signal, depending on how much oxygen is consumed in different brain areas. Its time course after stimulus presentation usually peaks around 5 s, but can last for up to 30 s. This is a much slower time scale than the millisecond temporal resolution of EEG and MEG. fMRI data are therefore usually not presented over time, but as one image that collapses brain activation over several seconds. There is still no clear model of how the electrophysiological sources of EEG and MEG relate to the metabolical fMRI signal. It cannot be generally assumed that activation hot spots in fMRI are also visible in EEG/MEG, or vice versa. Even if there is some correspondence, brain areas shown as activated in fMRI maps may activate at different time points in EEG/MEG. So far, some successful combinations of EEG/MEG with fMRI have been achieved for relatively simple perceptual processes. These are usually reflected in early EEG/MEG signals, and can be localised to a few focal brain areas.
In the example presented in the figure below, the combination of EEG and MEG for the estimation of early visually evoked responses is compared to fMRI data. Simple visual stimuli (socalled Gabor patches) were flashed in the four quadrants of the visual field (upper left, upper right, lower left, lower right). It is known that such stimuli activate characteristic retinotopic areas of limited extent (~ 1 cm2) in visual cortex in the occipital lobe (cortical area V1). Changing the stimulus location therefore changes the cortical response location in a predictable way. For example, visual stimuli presented in the left visual hemifield activate V1 in the right cortical hemisphere, and vice versa for right hemifield stimuli. Similarly, stimuli in the upper/lower visual field activate lower/upper parts of V1. This response pattern can then be compared across different localisation techniques.
The expected pattern of activation is clearly present in the fMRI data presented at the top of the figure below. This technique provides millimetre spatial resolution, but does not tell us whether this activation reflects early or late processes. EEG and MEG show a similar pattern around 74 ms after stimulus onset – this is about the time when visual information from the retina first reaches visual cortex. Localisations for EEG and MEG separately show approximately the same patterns, although EEG alone is quite far off in one case (left panel). The combination of EEG and MEG, however, nicely resembles the pattern obtained using fMRI. It confirms that this pattern reflects an early processing stage in visual cortex.
Note that the images for fMRI as well EEG/MEG are "tresholded", i.e. only activations above a certain value are shown. These represent the "tip of the iceberg", but the true spatial extent of activation can be difficult to determine. Especially for the case of EEG/MEG one has to keep in mind that spatial resolution is fundamentally limited, and that the estimated activations are always more blurred than the true underlying neural sources. Furthermore, there is no perfect match between fMRI and EEG/MEG. It is not clear whether this reflects the resolution limits of these methods, or fundamentally different physiological sources of the signals (electrical vs. metabolic). For this simple data set, the correspondence between the imaging techniques appears to be very high. However, for higher cognitive processes, the discrepancy may be much larger. A combination of these two imaging modalities should be considered with great care - an independent analysis and later comparison of fMRI and EEG/MEG may be more appropriate.
Multimodal.jpg

EEG:

Electroencephalography (EEG) is a non-invasive test for epilepsy during which several electrodes are placed on a patient's scalp to record electrical impulses from the brain (brain waves). It is sometimes called a brain wave test, used for testing patients with epilepsy, a brain tumor, a brain abscess, brain trauma, subdural hematoma, meningitis, encephalitis, stroke or congenital defects of the brain.
It is performed by using a device that measures the fluctuations and patterns in electrical processes within the brain.
                                               

 MRI:
MRI is a magnetic resonance imaging: the use of nuclear magnetic resonance of protons to produce proton density images. It is a type of diagnostic imaging that uses electromagnetic imaging and allows evaluation of tissues and fluid in addition to bone. This imaging providing two dimensional cuts through the body part being study which allows a physician to develop a three dimensional sense of the anatomy of that part. Often used to study the brain and spine as well as joints.
It is primarily used in medical imaging to visualise the structure and function of the body. It provides detailed images of the body in any plane.
MRI may detect certain lesions (diseased tissue) in the brain that cause seizures. MRI also helps specialists diagnose the causes of epilepsy and evaluate potential candidates for surgery.
Brain abnormalities may be subtle. Specialized MRI scanning may detect asymmetries (dissimilar corresponding parts on opposite sides that are normally alike) in the brain that indicate the side of the brain where seizures begin.

                                                 

FMRI:

Functional magnetic resonance imaging: a form of magnetic resonance imaging of the brain that registers blood flow to functioning areas of the brain.
It is a relatively new MRI technique that studies brain function. Using fMRI technology, scientists can determine which part of the central nervous system (CNS: brain and spinal cord) is active during a given task by tracking blood oxygen levels in the brain. Brain regions that are active require more oxygen. Oxygen is delivered by increasing the blood flow to these active brain regions. Scientists compare the differences in blood flow between a resting condition and an active condition, such as thinking, seeing, touching, or hearing, to find regions that are associated with one task and not the other.


PET:

Positron emission tomography (PET) is a nuclear medicine imaging technique which produces a three-dimensional image or picture of functional processes in the body. The system detects pairs of gamma rays emitted indirectly by a positron-emitting radionuclide (tracer), which is introduced into the body on a biologically active molecule. Images of tracer concentration in 3-dimensional or 4-dimensional space (the 4th dimension being time) within the body are then reconstructed by computer analysis. In modern scanners, this reconstruction is often accomplished with the aid of a CT X-ray scan performed on the patient during the same session, in the same machine.
A PET scan analyzes the amount of glucose, or sugar uptake by the brain. One would inject in a glucose substitute that has a label on it, and then take images of the brain during a period of time shortly after the seizure. The PET scan allows us to look at regions of the brain that are either hypermetabolic i.e. using a lot of sugar, or parts of the brain that are hypometabolic i.e. using very little sugar. It turns out that for Temporal Lobe Epilepsy, the temporal lobe shows hypometabolism.
                           

EEG vs fMRI and PET:

EEG has several strong points as a tool for exploring brain activity. EEG's can detect changes within a millisecond timeframe, excellent considering an action potential takes approximately 0.5-130 milliseconds to propagate across a single neuron, depending on the type of neuron. Other methods of looking at brain activity, such as PET and fMRI have time resolution between seconds and minutes. EEG measures the brain's electrical activity directly, while other methods record changes in blood flow (e.g., SPECT, fMRI) or metabolic activity (e.g., PET), which are indirect markers of brain electrical activity. EEG can be used simultaneously with fMRI so that high-temporal-resolution data can be recorded at the same time as high-spatial-resolution data, however, since the data derived from each occurs over a different time course, the data sets do not necessarily represent the exact same brain activity. There are technical difficulties associated with combining these two modalities, including the need to remove the MRI gradient artifact present during MRI acquisition and the ballistocardiographic artifact (resulting from the pulsatile motion of blood and tissue) from the EEG. Furthermore, currents can be induced in moving EEG electrode wires due to the magnetic field of the MRI


Electroencephalogram (EEG)

FUNDAMENTALS OF EEG MEASUREMENT
1. Introduction
Modern medicine applies variety of imaging techniques of the human body. The group of electrobiological measurements comprises items as electrocardiography (ECG, heart), electromyography (EMG, muscular contractions), electroencephalography (EEG, brain), magnetoencephalography (MEG, brain), electrogastrography (EGG, stomach), electrooptigraphy (EOG, eye dipole field). Imaging techniques based on different physical principles include computer tomography (CT), magnetic resonance imaging (MRI), functional MRI (fMRI), positron emission tomography (PET), and single photon emission computed tomography (SPECT).
Electroencephalography is a medical imaging technique that reads scalp electrical activity generated by brain structures. The electroencephalogram (EEG) is defined as electrical activity of an alternating type recorded from the scalp surface after being picked up by metal electrodes and conductive media [1].
The EEG measured directly from the cortical surface is called electrocortiogram while when using depth probes it is called electrogram. In this article, we will refer only to EEG measured from the head surface. Thus electroencephalographic reading is a completely non-invasive procedure that can be applied repeatedly to patients, normal adults, and children with virtually no risk or limitation.
When brain cells (neurons) are activated, local current flows are produced. EEG measures mostly the currents that flow during synaptic excitations of the dendrites of many pyramidal neurons in the cerebral cortex. Differences of electrical potentials are caused by summed postsynaptic graded potentials from pyramidal cells that create electrical dipoles between soma (body of neuron) and apical dendrites (neural branches). Brain electrical current consists mostly of Na+, K+, Ca++, and Cl- ions that are pumped through channels in neuron membranes in the direction governed by membrane potential [2]. The detailed microscopic picture is more sophisticated, including different types of synapses involving variety of neurotransmitters. Only large populations of active neurons can generate electrical activity recordable on the head surface. Between electrode and neuronal layers current penetrates through skin, skull and several other layers. Weak electrical signals detected by the scalp electrodes are massively amplified, and then displayed on paper or stored to computer memory [3]. Due to capability to reflect both the normal and abnormal electrical activity of the brain, EEG has been found to be a very powerful tool in the field of neurology and clinical neurophysiology. The human brain electric activity starts around the 17-23 week of prenatal development. It is assumed that at birth the full number of neural cells is already developed, roughly 1011 neurons [4]. This makes an average density of 104 neurons per cubic mm. Neurons are mutually connected into neural nets through synapses. Adults have about 500 trillion (5.1014) synapses. The number of synapses per one neuron with age increases, however the number of neurons with age decreases, thus the total number of synapses decreases with age too. From the anatomical point of view, the brain can be divided into three sections: cerebrum, cerebellum, and brain stem. The cerebrum consists of left and right hemisphere with highly convoluted surface layer called cerebral cortex. The cortex is a
1
Measurement in Biomedicine ● M. Teplan
dominant part of the central nervous system. The cerebrum obtains centres for movement initiation,
conscious awareness of sensation, complex analysis, and expression of emotions and behaviour. The
cerebellum coordinates voluntary movements of muscles and balance maintaining. The brain stem controls
respiration, heart regulation, biorythms, neurohormone and hormone secretion, etc.[5]. The highest
influence to EEG comes from electric activity of cerebral cortex due to its surface position.
There are some theoretical and practical differences between EEG and MEG. Although the MEG is
produced by the same electrical currents, it can provide complementary information to EEG [6].
2. EEG: History, brain waves, applications
History
During more than 100 years of its history, encephalography has undergone massive progress. The
existence of electrical currents in the brain was discovered in 1875 by an English physician Richard
Caton. Caton observed the EEG from the exposed brains of rabbits and monkeys. In 1924 Hans Berger, a
German neurologist, used his ordinary radio equipment to amplify the brain's electrical activity measured
on the human scalp. He announced that weak electric currents generated in the brain can be recorded
without opening the skull, and depicted graphically on a strip of paper. The activity that he observed
changed according to the functional status of the brain, such as in sleep, anaesthesia, lack of oxygen and in
certain neural diseases, such as in epilepsy. Berger laid the foundations for many of the present
applications of electroencephalography. He also used the word electroencephalogram as the first for
describing brain electric potentials in humans. He was right with his suggestion, that brain activity changes
in a consistent and recognizable way when the general status of the subject changes, as from relaxation to
alertness [7]. Later in 1934 Adrian and Matthews published the paper verifying concept of “human brain
waves” and identified regular oscillations around 10 to 12 Hz which they termed “alpha rhythm” [7].
Brain waves classification
For obtaining basic brain patterns of individuals, subjects are instructed to close their eyes and relax.
Brain patterns form wave shapes that are commonly sinusoidal. Usually, they are measured from peak to
peak and normally range from 0.5 to 100 μV in amplitude, which is about 100 times lower than ECG
signals. By means of Fourier transform power spectrum from the raw EEG signal is derived. In power
spectrum contribution of sine waves with different frequencies are visible. Although the spectrum is
continuous, ranging from 0 Hz up to one half of sampling frequency, the brain state of the individual may
make certain frequencies more dominant. Brain waves have been categorized into four basic groups
(Figure 1):
- beta (>13 Hz),
- alpha (8-13 Hz),
- theta (4-8 Hz),
- delta (0.5-4 Hz).




The best-known and most extensively studied rhythm of the human brain is the normal alpha rhythm. Alpha can be usually observed better in the posterior and occipital regions with typical amplitude about 50 μV (peak-peak). According to our experiences alpha was also significant between posterior and central regions in comparison to other regions. Alpha activity is induced by closing the eyes and by relaxation, and abolished by eye opening or alerting by any mechanism (thinking, calculating). Most of the people are remarkably sensitive to the phenomenon of “eye closing”, i.e. when they close their eyes their wave pattern significantly changes from beta into alpha waves. The precise origin of the alpha rhythm is still not known. Alpha waves are usually attributed to summated dendrite potentials. Evoked potentials (e.g. generated in brain stem) often consist of fibre potentials (axonal) and synaptic components [8].
EEG is sensitive to a continuum of states ranging from stress state, alertness to resting state, hypnosis, and sleep. During normal state of wakefulness with open eyes beta waves are dominant. In relaxation or drowsiness alpha activity rises and if sleep appears power of lower frequency bands increase. Sleep is generally divided into two broad types: nonrapid eye movement sleep (NREM) and REM sleep. NREM and REM occur in alternating cycles. NREM is further divided into stage I, stage II, stage III, stage IV. The last two stages corresponds to deeper sleep, where slow delta waves show higher proportions. With slower dominant frequencies responsiveness to stimuli decreases.
Various regions of the brain do not emit the same brain wave frequency simultaneously. An EEG signal between electrodes placed on the scalp consists of many waves with different characteristics. A large amount of data received from even one single EEG recording presents a difficulty for interpretation.
Individual’s brain wave patterns are unique. In some cases, it is possible to distinguish persons only according to their typical brain activity. For example, subjects who regard themselves as rational types or as holistic/intuitive types may demonstrate certain higher activity in their frontal left and frontal right hemisphere respectively.
Applications
The greatest advantage of EEG is speed. Complex patterns of neural activity can be recorded occurring within fractions of a second after a stimulus has been administered. EEG provides less spatial resolution compared to MRI and PET. Thus for better allocation within the brain, EEG images are often combined with MRI scans. EEG can determine the relative strengths and positions of electrical activity in different brain regions.
According to R. Bickford [8] research and clinical applications of the EEG in humans and animals are used to:
(1) monitor alertness, coma and brain death;
(2) locate areas of damage following head injury, stroke, tumour, etc.;
(3) test afferent pathways (by evoked potentials);
(4) monitor cognitive engagement (alpha rhythm);
(5) produce biofeedback situations, alpha, etc.;
(6) control anaesthesia depth (“servo anaesthesia”);
(7) investigate epilepsy and locate seizure origin;
(8) test epilepsy drug effects;
(9) assist in experimental cortical excision of epileptic focus;
(10) monitor human and animal brain development;
(11) test drugs for convulsive effects;
(12) investigate sleep disorder and physiology.
Symmetry of alpha activity within hemispheres can be monitored. In cases of restricted lesions such as tumour, hemorhage, and trombosis, it is usual for the cortex to generate lower frequencies. EEG signal distortion can be manifested by reduction in amplitude; decrease of dominant frequencies beyond the normal limit; production of spikes or special patterns. Epileptic conditions produce stimulation of the cortex and the appearance of high-voltage waves (up to 1000 μV) referred to as “spikes” or “spike and wave” [8]. EEG patterns have been shown to be modified by a wide range of variables, including biochemical, metabolic, circulatory, hormonal, neuroelectric, and behavioural factors [7]. By tracking
3
Measurement in Biomedicine ● M. Teplan
changes of electric activity during such drug abuse-related phenomena as euphoria and craving, brain areas and patterns of activity that mark these phenomena can be determined.
As the EEG procedure is non-invasive and painless, it is being widely used to study the brain organization of cognitive processes such as perception, memory, attention, language, and emotion in normal adults and children. For this purpose, the most useful application of EEG recording is the ERP (event related potential) technique.
Evoked potentials
Evoked potentials or event-related potentials (ERPs) are significant voltage fluctuations resulting from evoked neural activity. Evoked potential is initiated by an external or internal stimulus [8]. ERPs are suitable methodology for studying the aspects of cognitive processes of both normal and abnormal nature (neurological or psychiatric disorders [9]).
Mental operations, such as those involved in perception, selective attention, language processing, and memory, proceed over time ranges in the order of tens of milliseconds. Whereas PET and MRI can localize regions of activation during a given mental task, ERPs can help in defining the time course of these activations [10].
Amplitudes of ERP components are often much smaller than spontaneous EEG components, so they are not to be recognised from raw EEG trace. They are extracted from set of single recordings by digital averaging of epochs (recording periods) of EEG time-locked to repeated occurrences of sensory, cognitive, or motor events [11]. The spontaneous background EEG fluctuations, which are random relatively to time point when the stimuli occurred, are averaged out, leaving the event-related brain potentials. These electrical signals reflect only that activity which is consistently associated with the stimulus processing in a time-locked way. The ERP thus reflects, with high temporal resolution, the patterns of neuronal activity evoked by a stimulus.
Quantitative electroencephalography
Technological advances increased ability of encephalography to read brain activity data from the entire head simultaneously. Quantitative EEG (QEEG) applies multi channel measurements that can better determine spatial structures and localize areas with brain activity or abnormality. The results are often used for topographic brain mapping represented with colour maps in 2D and 3D to enhance visualization.
Brain computer interface
Brain computer interface (BCI) is a communication system that recognizes user’s command only from his or her brainwaves and reacts according to them. For this purpose PC or/and subject is trained. Simple task can consist of desired motion of an arrow displayed on the screen only through subject’s imaginary of the motion of his or her left or right hand. As the consequence of imaging process, certain characteristics of the brainwaves are raised and can be used for user’s command recognition, e.g. motor mu waves (brain waves of alpha range frequency associated with physical movements or intention to move) or certain ERPs.
EEG Biofeedback
So-called mindmaschines or brainmaschines are devices for induction of different mind states (e.g. relaxation, top performance) by entrainment of the brain waves into desired frequency bands by repetitive visual and audio stimuli. For making the training more effective, biofeedback methods were involved. Originally, changes in finger skin resistance or temperature were monitored. EEG biofeedback or neurofeedback uses EEG signal for feedback input. It is suggested that this learning procedure may help a subject to modify his or her brainwave activity. One of the methods involved in neurofeedback training is the so-called frequency following response. Changes in the functioning of the brain in desired way, e.g. increase in alpha activity, generates appropriate visual, audio, or tactile response. Thus, a person can be aware of the right direction of the training.
4
MEASUREMENT SCIENCE REVIEW,
Some researchers assume that subjects can improve their mental performance, normalize behaviour, and stabilize mood through the positive or negative feedback loop, while others are sceptical on these controversial issues. There are some findings indicating applications to certain range of conditions, as attention deficit disorder, depression, epilepsy, and alcoholism.
3. EEG recording techniques
Encephalographic measurements employ recording system consisting of
- electrodes with conductive media
- amplifiers with filters
- A/D converter
- recording device.
Electrodes read the signal from the head surface, amplifiers bring the microvolt signals into the range where they can be digitalized accurately, converter changes signals from analog to digital form, and personal computer (or other relevant device) stores and displays obtained data. A set of the equipment is shown in Figure 2.
Scalp recordings of neuronal activity in the brain, identified as the EEG, allow measurement of potential changes over time in basic electric circuit conducting between signal (active) electrode and reference electrode [12]. Extra third electrode, called ground electrode, is needed for getting differential voltage by subtracting the same voltages showing at active and reference points. Minimal configuration for mono-channel EEG measurement consists of one active electrode, one (or two specially linked together) reference and one ground electrode. The multi-channel configurations can comprise up to 128 or 256 active electrodes.
Figure




Recording electrodes
The EEG recording electrodes and their proper function are critical for acquiring appropriately high quality data for interpretation. Many types of electrodes exist, often with different characteristics. Basically there are following types of electrodes:
5
Measurement in Biomedicine ● M. Teplan
- disposable (gel-less, and pre-gelled types)
- reusable disc electrodes (gold, silver, stainless steel or tin)
- headbands and electrode caps
- saline-based electrodes
- needle electrodes
For multichannel montages, electrode caps are preferred, with number of electrodes installed on its surface (Figure 3). Commonly used scalp electrodes consist of Ag-AgCl disks, 1 to 3 mm in diameter, with long flexible leads that can be plugged into an amplifier [7]. AgCl electrodes can accurately record also very slow changes in potential [9]. Needle electrodes are used for long recordings and are invasively inserted under the scalp.
Skin preparation differs, generally cleaning of the skin surface from oil and brushing from dried parts is recommended. With disposable and disc electrodes, abrasive paste is used for slight skin abrasion. With cap systems, abutting needle at the end of injection is used for skin scraping, which can cause irritation, pain and infection. Especially when person’s EEG is measured repeatedly and cap is mounted for the same electrode points, there is a threat of certain pain and bleeding. That is why the right hygiene and safety protocol should be kept.
Using the silver-silver chloride electrodes, the space between the electrode and skin should be filled with conductive paste also helping to stick. With the cap systems, there is a small hole to inject conductive jelly. Conductive paste and conductive jelly serve as media to ensure lowering of contact impedance at electrode-skin interface.
In 1958, International Federation in Electroencephalography and Clinical Neurophysiology adopted standardisation for electrode placement called 10-20 electrode placement system [13]. This system standardized physical placement and designations of electrodes on the scalp. The head is divided into proportional distances from prominent skull landmarks (nasion, preauricular points, inion) to provide adequate coverage of all regions of the brain. Label 10-20 designates proportional distance in percents between ears and nose where points for electrodes are chosen. Electrode placements are labelled according adjacent brain areas: F (frontal), C (central), T (temporal), P (posterior), and O (occipital). The letters are accompanied by odd numbers at the left side of the head and with even numbers on the right side (Figure 4). Left and right side is considered by convention from point of view of a subject.
Figure 3. Electrode cap with electrodes placed after 10-20 electrode placement system.
6

As it is known from tomography different brain areas may be related to different functions of the brain. Each scalp electrode is located near certain brain centres, e.g. F7 is located near centres for rational activities, Fz near intentional and motivational centres, F8 close to sources of emotional impulses. Cortex around C3, C4, and Cz locations deals with sensory and motor functions. Locations near P3, P4, and Pz contribute to activity of perception and differentiation. Near T3 and T4 emotional processors are located, while at T5, T6 certain memory functions stand. Primary visual areas can be found bellow points O1 and O2. However the scalp electrodes may not reflect the particular areas of cortex, as the exact location of the active sources is still open problem due to limitations caused by the non-homogeneous properties of the skull, different orientation of the cortex sources, coherences between the sources, etc. [4].
High impedance can lead to distortions which can be difficult to separate from actual signal. It may allow inducing outside electric frequencies on the wires used or on the body. Impedance monitors are built in some commercially available EEG devices. In order to prevent signal distortions impedances at each electrode contact with the scalp should all be bellow 5 K Ohms, and balanced within 1 K Ohm of each other. Similar standard is required for clinical use of the EEG and for publication in most reputable journals. Practically, impedance of the whole circuit comprising two electrodes is measured, but built in impedance checks usually display results already divided by two. Control of all impedances is desirable also after finishing every single measurement.
Several different recording reference electrode placements are mentioned in the literature. Physical references can be chosen as vertex (Cz), linked-ears, linked-mastoids, ipsilateral-ear, contralateral-ear, C7 reference, bipolar references, and tip of the nose. Reference-free techniques are represented by common average reference, weighted average reference, and source derivation. Each technique has its own set of advantages and disadvantages. The choice of reference may produce topographic distortion if relatively electrically neutral area is not employed. Linking reference electrodes from two earlobes or mastoids reduces the likelihood of artificially inflating activity in one hemisphere. Nevertheless, the use of this method may drift away "effective" reference from the midline plane if the electrical resistance at each electrode differs [14]. Cz reference is advantageous when it is located in the middle among active electrodes, however for close points it makes poor resolution. Reference-free techniques do not suffer from problems associated with an actual physical reference. Referencing to linked ears and vertex (Cz) are predominant.
With modern instrumentation, the choice of a ground electrode plays no significant role in the measurement [15]. Forehead (Fpz) or ear location is preferred [16], but sometimes wrist or leg is also used. The combination of all active electrodes with reference and ground electrode compose channels. The general configuration is called montage.
Amplifiers and filters
The signals need to be amplified to make them compatible with devices such as displays, recorders, or A/D converters. Amplifiers adequate to measure these signals have to satisfy very specific requirements. They have to provide amplification selective to the physiological signal, reject superimposed noise and interference signals, and guarantee protection from damages through voltage and current surges for both patients and electronic equipment. The basic requirements that a biopotential amplifier has to satisfy are [17]:
- The physiological process to be monitored should not be influenced in any way by the amplifier.
- The measured signal should not be distorted.
- The amplifier should provide the best possible separation of signal and interferences.
- The amplifier has to offer protection of the patient from any hazard of electric shock.
- The amplifier itself has to be protected against damages that might result from high input voltages as
they occur during the application of defibrillators or electrosurgical instrumentation.
The input signal to the amplifier consists of five components:
The desired biopotential, undesired biopotentials, a power line interference signal of 50/60 Hz and its harmonics, interference signals generated by the tissue/electrode interface, and noise. Proper design of the amplifier provides rejection of a large portion of the signal interferences. The desired biopotential appears as the differential signal between the two input terminals of the differential amplifier [17].
The amplifier gain is the ratio of the output signal to the input signal. In order to provide optimum signal quality and adequate voltage level for further signal processing, the amplifier has to provide a gain of 100-100,000 [17] (the highest need not to be the best, combination of more parameters is involved, e.g. the range of the A/D converter, sampling rate, noise of the used elements) and needs to maintain the best possible signal-to-noise ratio. In order to decrease an impact of electrically noisy environment differential amplifiers must have high common-mode rejection ratios (at least 100 dB) and high input impedance (at least 100 M Ohms). The common-mode rejection ratio is the ratio of the gain of differential mode (wanted signal) over the gain of the common mode (original input signal between the inputs and ground).
Special electrically shielded rooms minimize the impact of urban electric background, in particular 50/60 Hz alternating current line noise. For usual medical purposes, shielded room is not necessary. For research purposes when maximal amount of information is desired, shielded room is used. Then amplifiers run on batteries and an optical cable leads to the PC standing outside from the shielded space. In addition to the optical cable, electrical/optical and optical/electrical converters are necessary. Usually information of interest lies bellow this line noise and we can use low-pass filters with cut-off bellow 50/60 Hz, or for keeping higher frequency bands a notch filter can be applied, that is able to reduce only a narrow band around 50/60 Hz (but distorts phases).
When computers are used as recording devices, channels of analog signal are repeatedly sampled at a fixed time interval (sampling interval), and each sample is converted into a digital representation by an analog- to-digital (A/D) converter. The A/D converter is interfaced to a computer system so that each sample can be saved in the computer’s memory. The resolution of the converter is determined by the smallest amplitude that can be sampled. This is obtained by dividing the voltage range of the A/D converter by 2 raised to the power of the number of bits of the A/D converter [7]. A/D converter usually uses minimally 12 bits (discerning 4,096 value levels). Ability to resolve 0.5 μV is recommended [18]. Sufficient sampling rate is required, at least double of the highest frequency component of our interest.
Analog (hardware) filters have to be integrated in the amplification unit. A high-pass filter is needed for reducing low frequencies coming from bioelectric flowing potentials (breathing, etc.), that remain in the signal after subtracting voltages toward ground electrode. Its cut-off frequency usually lies in the range of 0.1-0.7 Hz. To ensure that the signal is band limited, a low-pass filter with a cut-off frequency equal to the highest frequency of our interest [7] is used (in the range from 40 Hz up to less than one half of the sampling rate). Analog low-pass filters prevent distortion of the signal by interference effects with sampling rate, called aliasing, which would occur if frequencies greater than one half of the sampling rate survive without diminishing.
Once data are stored, digital filtering can be used. The strength of the analog filters is limited thus for displaying and processing of the signals further decreasing of DC components is usually needed. It is possible to choose from linear (FIR, IIR) filtering or novel non-linear filtering methods. The choice should be done according to the objectives put on the signal processing. Predominantly finite impulse response (FIR) filters are used which do not distort wave phases. The data points width typically range on the order of 1000 and one of the window function (Blackman, Hanning, Hamming, or rectangular) should be chosen. Filters should be designed in a way to influence useful signal properties minimally.
Before performing the final measurements the whole EEG system should be tested. Inter-channel calibrations with known wave signal parameters should not display significant discrepancies. The output noise (referred to input) consists mainly from the noise caused by the analog amplifier circuitry and by A/D converter circuitry. Noise value should be consistent with manufacturer information, about 0.3-2 μV pp. (range from negative peak to positive peak) but this value depends on the way of noise estimation and on the system configuration (low-pass filter, sampling rate, choice of circuitry). The noise can be determined by connecting the inputs of the amplifier together, or abased them into a salty solution, or "short-circuiting" the inputs, and then measuring the output of the amplifier. The number of useful information bits can be counted as a power of two from the ratio of average EEG signal amplitude over the noise amplitude (e.g. 50μV/1μV results in over 5 bits).
One of the limitations of recordings is due to storage requirements. For example, 1 hour of eight channels 14-bit signal sampled with 500 Hz occupies 200 MB of the memory. There exist portable recording systems used for longer monitoring of a subject without limiting movement of a person. Some of the commercial EEG recording systems comes from following suppliers: Lexicor, Electrical geodesics, Biosemi, NeuroScan, Sigma Medizin, Contact Precision Instruments, Stellate, Thought Technology, Xltek.
Artefacts
Among basic evaluation of the EEG traces belongs scanning for signal distortions called artefacts. Usually it is a sequence with higher amplitude and different shape in comparison to signal sequences that doesn’t suffer by any large contamination. The artefact in the recorded EEG may be either patient-related or technical. Patient-related artefacts are unwanted physiological signals that may significantly disturb the EEG. Technical artefacts, such as AC power line noise, can be decreased by decreasing electrode impedance and by shorter electrode wires. The most common EEG artefact sources can be classified in following way:
Patient related:
- any minor body movements
- EMG
- ECG (pulse, pace-maker)
- eye movements
- sweating
Technical:
- 50/60 Hz
- impedance fluctuation
- cable movements
- broken wire contacts
- too much electrode paste/jelly or dried pieces
- low battery
Excluding the artefact segments from the EEG traces can be managed by the trained experts or automatically. For better discrimination of different physiological artefacts, additional electrodes for monitoring eye movement, ECG, and muscle activity may be important.

Notes on experimental set-up
For data acquisition of suitable quality, a choice of the right representative group of volunteers with appropriate spectrum of demographic data and other parameters should be considered. The position of subjects during the EEG measurements should be comfortable enough to avoid unwanted activities, a lying position diminishes the occurence of some artefacts caused by feeble motion, on the other hand it may lead towards sleeping, especially when a darkened noiseless room is designed. Keeping of the same conditions and instructions for the subjects for the whole length of the experiment period is desired.
Conclusion
Electroencehalography belongs to electrobiological imaging tools widely used in medical and research areas. EEG measures changes in electric potentials caused by a large number of electric dipoles formed during neural excitations. EEG signal consists of different brain waves reflecting brain electrical activity according to electrode placements and functioning in the adjacent brain regions. For using EEG techniques, the following recording system components are necessary:
- Electrode cap with conductive jelly or Ag-AgCl disc electrodes with conductive paste.
- Amplifiers with overall amplification gain between 100-100,000, with input impedances at least 100 M Ohms, and common-mode rejection ratio at least 100 dB.
- Analog filters integrated in the unit with high pass filter with cut-off frequency in the range of 0.1-0.7 Hz and low pass filter with cut-off frequency less than one half of the sampling rate. In fact, frequencies above 50 Hz are rarely involved as they contribute negligibly to power spectrum of EEG.
- At least 12 bit A/D converter with accuracy lower than overall noise (0.3-2 μV pp.), and sampling frequency usually between 128 – 1024 Hz.
- Sufficiently quick PC for taking over data for recording and eventually for online analysis, with adequate volume of hard disc.
- Digital high pass FIR filter with similar cut-off frequency as analog high pass.
The general quality of recording equipment depends on the right combination of the mentioned parameters. Before further data processing, raw EEG signal should be checked for artefacts.
 

Wednesday, September 25, 2013

Some Latest Technologies



Some Upcoming Technologies.............
1.     WiSee
2.     Self-assembling copter
3.     ATLAS (DARPA)
4.     V-bat drone
5.     Calligraphy Robot
6.     Mech Warfare
7.     The V Motion Project
8.     Qumarion mannequin
9.     Artificial Cockroach’s sensor network
10.                        Interactive plants
11.                        Magnetoreceptor cell
12.                        Color-changing contact lenses which reveal glucose concentration
13.                        Ultrasound surgery, healing without cuts
14.                        iCHELL Inorganic cells imitate properties of Living Organic Cells
15.                        Phonebloks
16.                        Project LOON
17.                        LuminAR
18.                        Visible Light Communication
19.                        The Centrifuge Brain Project
20.                        Fire Control using Sound Waves
21.                        Perceptual Motion
22.                        The Water and Sound Experiment
23.                        Ambient Backscattering
24.                        The Hear Colors Experiment
25.                        Cyborg:   Niel Harmission (Chromotopsia)
26.                        Position Correcting Router
27.                        Segway Wheelchair
28.                        HUD Contact Lenses
29.                        Panasonic Evolta batteries
30.                        Stalker UAS
31.                        Recycling ROBOT
32.                        AQUA Lungs (Oceanwings)
33.                        Hunter-Killer Aerial Terminator
34.                        Spaceplane
35.                        SpaceGlasses
36.                        Virtuix Omni Rift
37.                        Muscle Propelled Force FeedBack system
38.                        EGO SmartMouse
39.                        EDV-01 Disaster Response Unit
40.                        Stealth Barcodes
41.                        WI-Vi
42.                        Navy Laser Guns
43.                        X-51A Waverider
44.                        Moving Sidewalks
45.                        Touchy: A Human Camera
46.                        Brymill Cry-Ac 3
47.                        Volvo flywheel KERS
48.                        Thermite
49.                        Phone Scent
50.                        4D Ultrasound Hologram
51.                        ARGUS 1.8 : A Giga Pixel Drone Camera
52.                        HDR Vision Technology
53.                        Eye Ring
54.                        Visualization of video frames at speed of light
55.                        Light Field Camera
56.                        Eyeborg
57.                        Light Sabers